Design and Optimization Methodology of Sub-dermal Electroencephalography Dry Spike-Array Electrode
The purpose of this methodology is to meet the design specifications for portable long-term EEG recording and optimize the electrical performance of the electrodes by maximizing the electrode-skin contact surface area, while fulfilling design constraints including mechanical, physiological and economical limitations. This was followed by proposing a low cost fabrication technique to implement the electrodes. The proposed electrode design has a potential impact in enhancing the performance of the current recording systems, and also suits portable monitoring and long term recording devices. The design process was aided by using a software design and optimization tool, which was specifically created for this application.
The application conditions added challenges to the electrode design in order to meet the required performance requirements. On the other hand, the required design specifications are not fulfilled in the current electrode technologies which are designed and customized only for short term clinical recordings.
The electrode theory of application was verified using an experimental setup for an electrochemical cell, but the overall performance including measuring the electrode impedance is awaiting a clinical trial.